Abstract


The evaluation of uncertainty measurement in the determination of Fe content in powdered tonic food drink using graphite furnace atomic absorption spectrometry was carried out. The specification of measurand, source of uncertainty, standard uncertainty, combined uncertainty and expanded uncertainty from this measurement were evaluated and accounted. The measurement result showed that the Fe content in powdered tonic food drink sample was 569.32 µg/5g, with the expanded uncertainty measurement ± 178.20 µg/5g (coverage factor, k = 2, at confidende level 95%). The calibration curve gave the major contribution to the uncertainty of the final results.


Keywords


uncertainty; powdered tonic food drink; iron (Fe); graphite furnace AAS

Full Text:

Full Text PDF


References

[1] Ellison, S.L.R., Rosslein, M., and Williams, A., 2000, EURACHEM/CITAC Guide CG 4, Quantifying Uncertainty in Analytical Measurement, Second Edition,

[2] Singh, S., Arora, M., and Ramkishore, 2007, MAPAN-J. Metrology. Soc. of India, 22, 4, 263-268.

[3] National Standard Method, Uncertainty of Measurement in Testing, Standards Unit, Evaluations and Standards Laboratory Centre for Infections, 2005

[4] Kirkup, L., 2007, Calculating and Expressing Uncertainty in Measurement, University of Technology Sydney.

[5] Portugal, T.R., Udarbe, M.A., Rosales, J.D.C., Bisonaya, M.V., and Soliven, J.O., 2008, FNRI-DOST Proficiency Test Report No.02, Round 2 Powdered Tonic Food Drink: Proximates (Moisture, Fat, Protein, and Ash) and Minerals (Iron, Calcium, and Sodium), Food Nutrition Research Institue, Department of Science and Technology, Philippines.

[6] Horwitz, W., 2000, Official Methods of Analysis of AOAC International: Food Composition, Additives, Natural Contaminants, 17th edition, vol 2, AOAC International, Maryland-USA

[7] Chui, S.Q.H., 2007, J. Braz. Chem. Soc., 18, 2, 424-430.

[8] ISO Guide to The Expression of Uncertainty of Measurement, International Organization for Standardization, Geneva Switzerland, 1993.



DOI: https://doi.org/10.22146/ijc.21538

Article Metrics

Abstract views : 3585 | views : 3867


Copyright (c) 2010 Indonesian Journal of Chemistry

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

 


Indonesian Journal of Chemistry (ISSN 1411-9420 /e-ISSN 2460-1578) - Chemistry Department, Universitas Gadjah Mada, Indonesia.

Web
Analytics View The Statistics of Indones. J. Chem.