The Utilization of Nitrogen Gas as a Carrier Gas in the Determination of Hg Ions Using Cold Vapor-Atomic Absorption Spectrophotometer (CV-AAS)

https://doi.org/10.22146/ijc.23092

Aman Sentosa Panggabean(1*), Subur P Pasaribu(2), Farida Kristiana(3)

(1) Department of Chemistry, Faculty of Mathematics and Natural Sciences, Mulawarman University, Jl. Barong Tongkok No.4 Kampus Gn. Kelua Samarinda 75123, Indonesia
(2) Department of Chemistry, Faculty of Mathematics and Natural Sciences, Mulawarman University, Jl. Barong Tongkok No.4 Kampus Gn. Kelua Samarinda 75123, Indonesia
(3) Department of Chemistry, Faculty of Mathematics and Natural Sciences, Mulawarman University, Jl. Barong Tongkok No.4 Kampus Gn. Kelua Samarinda 75123, Indonesia
(*) Corresponding Author

Abstract


The research about utilization of nitrogen gas as a carrier gas in the determination of Hg ions by using Cold Vapor-Atomic Absorption Spectrophotometer (CV-AAS) method has been conducted. To optimize the measurement results, several parameters that affect hydride generator have been studied. Some specified important parameters are SnCl2 concentration as reductant, acid concentration, and the analytical performance such as repeatability and reproducibility (% RSD), linearity (r), limits of detection (LOD) and limits of quantitation (LOQ), and accuracy have been studied. The results of the research showed that the nitrogen gas can be used instead of argon gas as a carrier gas. It was shown by the repeatability values as % RSD < 2/3 CV Horwitz values, the LOD was 0.0338 µg/L and the LOQ was 0.0838 µg/L. The accuracy of this method was well shown by a recovery percentage was 102.24%. Based on the result of this research, the nitrogen gas can be used as a carrier gas for the determination of Hg ions by using the CV-AAS method with valid results.

Keywords


Hg; carrier gas; nitrogen; CV-AAS

Full Text:

Full Text PDF


References

[1] Li, P., Feng, X., and Qiu, G., 2010, Methylmercury exposure and health effects from rice and fish consumption: A review, Int. J. Environ. Res. Public Health, 7 (6), 2666–2691.

[2] Chen, Y., Dong, X., Dai, Y., Hu, Q., and Yu, H., 2008, Determination of trace mercury in Chinese herbal medicine by cold vapour generation atomic fluorescence spectrometry, Asian J. Chem., 20 (6), 4639–4646.

[3] Tessier, A., and Turner, D.R., 1995, Metal Speciation and Bioavailability in Aquatic Systems, IUPAC Series on Analytical and Physical Chemistry of Environmental System, Vol. 3, John Wiley and Sons Ltd, Chicester, 679.

[4] Shpirt, M.Ya., and Punanova, S.A., 2011, Accumulation of mercury in petroleum, coal, and their conversion products, Solid Fuel Chem., 45, 330–336.

[5] do Nascimento, F.H., and Masini, J.C., 2012, Complexation of Hg(II) by humic acid studied by square wave stripping voltammetry at screen-printed gold electrodes, Talanta, 100, 57–63.

[6] Vicentino, P.O., Brum, D.M., and Cassella, R.J., 2015, Development of a method for total Hg determination in oil samples by cold vapor atomic absorption spectrometry after its extraction induced by emulsion breaking, Talanta, 132, 733–738.

[7] Oreste, E.Q., de Jesus, A., de Oliveira, R.M., da Silva, M.M., Vieira, M.A., and Ribeiro, A.S., 2013, New design of cold finger for sample preparation in open system: Determination of Hg in biological samples by CV-AAS, Microchem. J., 109, 5–9.

[8] dos Santos, W.N.L., Dias, F.S., Reboucas, M.V., Pereira, M.G., Lemos, V.A., and Teixeira, L.S.G., 2006, Mercury determination in petroleum products by electrothermal atomic absorption spectrometry after in situ preconcentration using multiple injections, J. Anal. At. Spectrom., 21 (11), 1327–1330.

[9] Uddin, R., Al-Fahad, M.A., Al-Rashwan, A.K., and Al-Qarni, M.A., 2013, A simple extraction procedure for determination of total mercury in crude oil, Environ. Monit. Assess., 185 (5), 3681–3685.

[10] Chen, F.Y., and Jiang, S.J., 2009, Determination of Hg and Pb in fuels by inductively coupled plasma mass spectrometry using flow injection chemical vapor generation, Anal. Sci., 25 (12), 1471–1476.

[11] Saint'Pierre, T.D., Rocha, R.C.C., and Duyck, C.B., 2013, Determination of Hg in water associate to crude oil production by electrothermal vaporization inductively coupled plasma mass spectrometry, Microchem. J., 109, 41–45.

[12] Caiminagua, A., Fernández, L., Romero, H., Lapo, B., and Alvarado, J., 2015, Electrochemical generation of arsenic volatile species using a gold/mercury amalgam cathode. Determination of arsenic by atomic absorption spectrometry, Anal. Chem. Res., 3, 82–88.

[13] Kumar, A.R., and Riyazuddin, P., 2005, Mechanism of volatile hydride formation and their atomization in hydride generation atomic absorption spectrometry, Anal. Sci., 21 (12), 1401–1410.

[14] Silva, L.O.B., Leao, D.J., dos Santos, D.C., Matos, G.D., de Andrade, J.B., and Ferreira, S.L.C., 2014, Determination of copper in airborne particulate matter using slurry sampling and chemical vapor generation atomic absorption spectrometry, Talanta, 127, 140–145.

[15] Thompson, K.C., and Thomerson, D.R., 1974, Atomic absorption studies on the determination of antimony, arsenic, bismuth, germanium, lead, selenium, tellurium and tin by utilising the generation of covalent hydrides, Analyst, 99 (1182), 595–601.

[16] Nakahara, T., 2005, Development of gas-phase sample-introduction techniques for analytical atomic spectrometry, Anal. Sci., 21 (5), 477–484.

[17] Panggabean, A.S., Pasaribu, S.P., Amran, M.B., and Buchari, B., 2013, Gas-liquid separator integrated to HG-GFAAS method for determination of tin at trace levels in the water samples, Eurasian J. Anal. Chem., 8 (1), 17–27.

[18] Dantas, H.V., Barbosa, M.F., Moreira, P.N.T., Galvão, R.K.H., and Araújo, M.C.U., 2015, An automatic system for accurate preparation of gas mixtures, Microchem. J., 119, 123–127.

[19] D'Ulivo, A., 2004, Chemical vapor generation by tetrahydroborate(III) and other borane complexes in aqueous media: A critical discussion of fundamental processes and mechanisms involved in reagent decomposition and hydride formation, Spectrochim, Acta, Part B, 59 (6), 793–825.

[20] Ritschdorff, E.T., Fitzgerald, N., Mclaughlin, R.G.J., and Brindle, I.D., 2005, The use of a modified Multimode Sample Introduction System for the simple and rapid determination of cadmium by chemical vapour generation atomic absorption spectrometry, Spectrochim. Acta, Part B, 60 (1), 139–143.

[21] Miller, J.C., and Miller, J.N., 1991, Statistika Untuk Kimia Analitik, 2nd ed., ITB Press, Bandung.

[22] Panggabean, A.S., Pasaribu, S.P., Bohari, B., and Nurhasanah, H., 2014, Preconcentration of Chromium(VI) at trace levels using acid alumina resin with column method, Indones. J. Chem., 14 (1), 51–56.



DOI: https://doi.org/10.22146/ijc.23092

Article Metrics

Abstract views : 3535 | views : 3902


Copyright (c) 2018 Indonesian Journal of Chemistry

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.

 


Indonesian Journal of Chemistry (ISSN 1411-9420 /e-ISSN 2460-1578) - Chemistry Department, Universitas Gadjah Mada, Indonesia.

Web
Analytics View The Statistics of Indones. J. Chem.